What do I do if …? • 77
English
CAUTION!
Personal injury
If you notice any faults or deviations during the
function check which cannot be resolved with
the aid of the section entitled “What do I do if
...” in this document, you may not use the
device.
Have the device repaired by seca service or
by an authorized service partner.
Follow the section entitled “Servicing/
recalibration” in this document.
10.WHAT DO I DO IF …?
Fault Cause/Remedy
… no weight is dis-
played when scale is
loaded?
The scale is not connected to the power supply.
- Check if the scale is switched on.
- Check whether the battery block is inserted
and that the plug contacts are properly con-
nected.
… 0.000 does not appear
before weighing?
A load was placed on the scale before it was
switched on.
- Remove load.
- Switch scale off and back on again.
… a segment lights up
continuously or not at
all?
The relevant segment is faulty.
- Please notify maintenance service.
is displayed?
Battery block power is running down.
- Charge battery block.
… the display “bAtt”
appears?
Battery block discharged.
- Charge battery block.
… the display “StOP”
appears?
The maximum load has been exceeded.
- Remove load.
… the display “tEMP”
appears?
The ambient temperature of the scale is too high
or too low.
- Set up scale in an ambient temperature
between +10 °C and +40 °C.
- Wait approx. 15 minutes until scale has
adjusted to ambient temperature.